Figure 2 Product components
1 TU5200
2 StablCal kit, sealed vials with RFID (10, 20 and
600 NTU)
3 Sample vials
Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an
altitude higher than 3100 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.
Installation guidelines
Install the instrument:
• On a level surface
• In a clean, dry, well ventilated, temperature controlled location
• In a location with minimum vibrations that has no direct exposure to sunlight
• In a location where there is sufficient clearance around it to make connections and to do
maintenance tasks
• In a location where the power button and power cord are visible and easily accessible
Connect to external devices (optional)
The instrument has three USB 1.1 ports and one Ethernet port. Refer to
4 Power supply with adapters
5 Dust cover
6 Vial stand
C A U T I O N
Figure 1
on page 8.
English 9