Keep the battery compartment dry when the battery is replaced. If water gets into the battery
compartment, remove the battery, pour the water out and fully dry the compartment with a towel. Let
the battery compartment air dry fully before use to prevent corrosion.
Note: The internal clock battery is not user-replaceable without authorization from the manufacturer.
Examine the O-rings and sealing gasket
1. At regular intervals, examine the O-rings and sealing gasket in the mooring cap for unwanted
material, wear or damage. Refer to
2. Remove any unwanted material from the O-rings and mooring cap.
3. Replace any O-rings with visible damage or wear (e.g., cracks or missing pieces). Replace with
the O-rings from the manufacturer.
4. Apply silicone grease to the O-rings to prevent them from sticking and being pulled out of
position.
5. Replace the mooring cap when damage or wear is visible on the sealing gasket.
Figure 9 Sealing gasket location
1 Mooring cap sealing gasket
Figure 9
and
Figure
10.