Use
3.7
Possible results of the analysis
Following the completion of
the
analysis
the Analyzer
shows the results of the
analysis on the display. The
results
are
displayed
specifically for each tested
substance.
Four results of the analysis are possible:
Negative
The sample does not contain the wanted substance/group of
substances or only in a concentration well below the limiting
value of this detection reaction (see "Intended use" on page 7).
Positive
The sample does contain the wanted substance/group of
substances or cross-reactive substances (e.g. metabolites of
the wanted substance) in a concentration above the limiting
value of this detection reaction (see "Intended use" on page 7).
Invalid
Each detection reaction has an integrated process control
which is active during evaluation. Only if a detection reaction
has been executed correctly will the Analyzer take this specific
detection reaction into account. In case of errors in the reaction
process the affected detection reaction will not be evaluated to
exclude wrong results and is marked as "invalid". Only affected
detection reactions are marked "invalid", not affected detection
reactions are further evaluated and displayed as "positive/
negative". Repeat the analysis if you require a result for the
unevaluated substances/group of substances.
12
3.8
Switching Analyzer off
1. Keep
OK
3 seconds. The power supply cable should remain
connected to charge the battery, until the unit is returned to
the packaging, e.g. for transporting.
When switched off, the Analyzer will emit two short beeps
when the power pack is disconnected.
2. Follow the instructions on battery care on page 25.
In order to protect the optical components of the Analyzer
against dust, the door of the Analyzer should remain closed
07323950_en_ca.eps
(unless you are inserting or removing test kits).
-key pressed. The unit will switch off after approx.
Dräger DrugTest 5000 Test System