vI.2 - Sterilization
• Only sterilize MDs or accessories that have been cleaned.
• We strongly recommend that the sterilization procedure comply with
Standard EN 13060, class B. Any other type should be prohibited.
• After selecting the most appropriate bag sizes, put each MD and /or
accessories into an individual sterilization bag that complies with current
standards.
• Allow enough space between bags and do not overload the autoclave.
g
• The MDs and their accessories can be sterilized at 135°C for a
minimum of 3 minutes (sterilization time).
• Ensure that the device is perfectly dry at the end of the cycle.
vII.1 – choosing the post
Place the calibrating film on the periapical X-ray.
The calibrating film is full scale, be sure to check the scale of the X-ray.
Select the obturation product best suited for the case (length – diameter).
vII.2 – hygiene
All MDs used intra-orally must be cleaned, disinfected and sterilized (See
§ VI).
vII.3 – Preparing the canal
Use the Gates drill to eliminate the remaining gutta percha in the canal.
Progressively widen the canal by using drills of increasing diameter. The
last drill used should be the same size as the MD to be placed.
E
• Ensure that sufficient dentinal wall remains.
• Use the stops and the ruler to adjust the reaming depth. The
stops are fixed on the drill and provide a visual indication of the
depth to drill.
• Check that the colour codes for both the accessories (drills/rea-
mers or keys) and the MDs are the same ones.