Use a dam to avoid any slipping of the MD or its accessories (drill-reamer-
burr-key) within the mouth.
The practitioner is responsible for selecting the size, shape and material
for the post.
There are risks of interactions between the MD and the local environment.
Only the practitioner can assess what type of material has to be used
(titanium or stainless steel).
To avoid breaking the castable posts in the mouth, never fit them intra-
orally.
Drills/reamers must be replaced every 20 to 40 uses.
See summary tables
Posts :
Posts :
Screw-Posts: Titanium EUROPOST
F
The MD and all its accessories are supplied non-sterile.
Prior to use, the MD must be cleaned, decontaminated and sterilized (See
§ VI).
12
Iv.DEScRIPTIf DES DM
PIVOT MASTER cylinder-pointed posts
PIVOT MASTER conical posts
PIVOT MASTER cylinder-pointed locks
PIVOT MASTER conical locks
RVS EXCEL POST
TVS EXCEL POST
Stainless steel EUROPOST
v.MD START UP PROcEDURE