7.6.1
Determination of the average piece weight by weighing
Set reference
Reset balance to zero or tare the empty weighing container if necessary.
Place on the weighing plate a known number (e.g. 10 items) of individual pieces
as a reference.
Wait for the stability display, than enter the number of individual items via the
numeric keypad.
Acknowledge with
The balance determines the average piece weight.
Count the items
Tare if necessary, place weighing good and read off the number of items.
Delete reference
Press
, the average unit weight will be deleted.
24
.