Temperature setting
Temperature control:
Temperature regulation:
Error codes on the display
Display code
Description
- o -
The door is open
A1
Door alarm "dAd" from LAL and/or EAL has been activated
A2
Local upper alarm LHL is or has been avtivated
A4
External upper alarm EHL is or has been activated
F1
Cabinet sensor error. In the meantime the cabinet itself will maintain
the set temperature by the memory of the controller. Temperature
stability will be affected.
Service assistance is required.
F2
Evaporator sensor error. Service assistance is required.
The cabinet will keep running until the error has been mended.
F3
Condenser sensor error. The cabinet will keep running, until the error
has been mended. Service assistance is required.
F7
Indicates that the condenser temperature is too high. The cause might
be a clogged condenser, or too high ambient temperature.
If the condenser or air filter needs cleaning, the cabinet must be
disconnected at the mains power. Cleaning of the condenser is done
with a brush or a vacuum cleaner.
The air filter can be removed and cleaned in a dishwasher at max.
50C.
If the ambient temperature is too high, the placement of the cabinet
might be wrong, and an alternative place should be found. Ventilation
might help.
If this does not help, request service assistance.
765041572
P
Press
to see the cabinet temperature setting; the display will show
the set temperature.
P
Keep
pressed. At the same time, press
-
+
Each time
or
degree. When the display shows the desired temperature, let go of the
two keys and the setting has been made.
Rev. 002
is pressed the temperature will change one
-
+
or
.
40