Error Message Hb 10 Digital/Control - IKA HB 10 basic Mode D'emploi

Table des Matières

Publicité

Error message HB 10 digital/control

Any malfunctions during operation will be identified by an error message on the display (H), see fig. 6.
Proceed as follows in such cases:
☞ Switch off device using the main switch (B)
☞ Carry out corrective measures
☞ Restart device
Error message
E 2
E 3
E 9
E 21
E 23
E 24
E 25
E 26
E 27
E 28
E 29
E 30
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
032009
Cause
- In remote operation (PC), no communication
between RV 10 and HB 10
- IR connection to RV 10 disconnected
- Temperature inside the device is too high
- Fault in storing the safety circuit set temperatures-
- Memory chip (EPROM) defective
- Safety relay does not open
- Adjustable safety circuit defective
- Set safety temperature exceeded
- The heater control circuit switch (TRIAC) is faulty
The heater or the supply line is disconnected
- Dry running
- Error in calibration
- Breakage of the controller sensor
- Short-circuit in the safety sensor
- Short-circuit in the controller sensor
Effect
Correction
Heating off
- Switch off the device
- Clean the IR interface or
remove objects from around the IR interface
- Switch on the device
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device
Heating off
- Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device again
- Check the safety temperature setting
Heating off
- Switch off the device and switch on again
Heating off
- Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
23