5 Memory / 6 Miscellaneous
the M-button
measurement. Repeatedly pressing the M-button
tive values measured previously. Press the
more actual stored results. The unit switches off automatically after
approx. one minute if no buttons are pressed. The memory recall mode
may be exited at any time by pressing the START/STOP button
The oldest value is deleted if 500 measured values have been stored in
the memory and a new value is saved.
5.2 Deleting memory
Press M-button
If you only want to delete the last measured value, press and hold the S-
button
for approx. 7 seconds. The last measured value will flash two
times and will then be deleted. The display shows "dEL". Press START/
STOP
, to exit the delete mode. If you want to delete all stored
values in memory for a user, press and hold M-button
simultaneously for approx. 10 seconds. The display shows "dEL ALL".
S-button
Now press the
"dEL OK" - now all stored values for this user are deleted and the unit will
switch off automatically. If you want to abort the deletion process, press
START / STOP
switches itself off without deleting any value.

6 Miscellaneous

6.1 Error messages and error remedying
Error indicators:
Symbol
Cause
+ Lo Low battery
Bluetooth transfer of
+ Err
data not possible
E1
Cuff not properly
fitted
E2
Cuff too tight
E3
Failure during
inflation
E10/E11
Movement or speaking
during measurement or
pulse too weak
26
again to display the values from the previously stored
to enter the memory recall mode.
to confirm the deletion. The display shows
while the display shows "dEL ALL". The unit then
displays the respec-
S-button
to browse to the
Correction
The built-in battery is too low or
empty. Charge the battery.
Check the software and if Blue-
tooth is properly configured and
activated on the receiving units.
Wrap the cuff properly.
Remeasure in correct way.
Wrap the cuff properly.
Remeasure in correct way.
Wrap the cuff properly.
Remeasure in correct way.
Rest 30 minutes and
in correct way. Do not speak or
move during the measurement.
.
i
AND S-button
remeasure