Over-limit setting range: 25bpm~250bpm
Default setting limit: High -- 120bpm, Low -- 50bpm
E. Perfusion Index Display
Range: 0.2%~20%
F. Operating Environment
Operating Temperature: 5°C ~40°C
Operating Humidity: 30%~80%
Atmospheric pressure: 70kPa~106kPa
Note: portable and mobile RF communications equipment may affect the
performance of the Oximeter.
G. Data Update
8 beats moving average for both SpO
and pulse rate readings
2
H. Data Storage
Recording SpO
and pulse rate data every 1/2/4/8 second(s), up to 384–hour
2
records can be stored.
I. Low Perfusion Performance
The accuracy of SpO
and PR measurement still meet the precision described
2
above when the modulation amplitude is as low as 0.5%.
J. Resistance to interference of surrounding light:
The difference between the SpO
value measured in the condition of indoor
2
natural light and that of darkroom is less than ±1%.
K. Dimensions: 145 mm (L) × 74 mm (W) × 29 mm (H)
Net Weight: 210g (including batteries)
L. Classification
Type of protection against electric shock: Internally powered equipment
Degree of protection: Type BF applied parts.
Degree of protection against harmful ingress of liquids: IP22. The device is not
intended for use in the environment with rich oxygen or ignitable gas.
Mode of operation: Continuous operation.
Electro-Magnetic Compatibility: Group I, Class B
6 Over-limit Indication
6.1 Limit settings
- SpO
low limit setting range: 50% ~ 99%.
2
- Pulse Rate limits setting range:
High: 100bpm--240bpm Low: 30bpm - 99bpm
During the measurement, if the measured value exceeds the preset value,
the alert beeping sound will be activated, the value that is over-limit will blink
at the same time.