Fault – Cause – Remedy
Message
Malfunction
Observe Service
Interval
Pump error
Unknown chip type
Remove chip 0
Other faults
Fault
Sliding switch is in
position 0, but the chip is
not ejected from the chip
outlet.
60
Cause
RAM/ROM/EEPROM/IRQ
fault.
Analyzer has detected an
internal fault.
Analyzer has performed
approx. 500 measurements.
Pump capacity inadequate.
Pump system faulty.
Chip type cannot be
processed.
Possible cause
Screw securing the Remote
System or cover of the
compartment protrudes into
the chip channel.
Remedy
Switch off analyzer:
move sliding switch to position 0,
wait at least 1 minute then switch
on again.
Call DrägerService if the fault
cannot be remedied in this way.
Analyzer must be inspected and
serviced by duly trained specialist
personnel.
Ensure that intake channel in face
end of analyzer is clear.
Call DrägerService.
Call DrägerService.
Remedy
Back off the screw slightly.
Dräger CMS Analyzer