technical specifications
Caring for your
ACTIVATOR
Your
should not be allowed to get wet or be left in excessive sunlight. It may be
ACTIVATOR
cleaned regularly using a soft cloth, lightly dampened in soapy water. Do not allow the
interior of the
to become wet. Do not use detergents, alcohol, spray aerosols or
ACTIVATOR
strong solvents on the
.
ACTIVATOR
Access to the interior of the
ACTIVATOR
If your
is damaged, you should not use it, but should return it to
ACTIVATOR
retailer for replacement or repair. Repairs, service and modifications may not be carried
out by anyone other than qualified service personnel authorized by
: Under no circumstance should anything other than the correct type of batteries
IMPORTANT
- (rechargeable batteries 3.6V NiMH) be used with your
from the Careline.
Disposing of the GelPads and batteries
Used GelPads and batteries must never be disposed of in a fire but in accordance with
national or state laws governing the disposal of such items.
: The signal gradually increases to a peak intensity level at the start of the contraction
NOTE
phase (i.e. ramp up) and gradually decreases to nothing at the start of the relaxation phase
(i.e. ramp down). When the
(i.e. ramp down). When the
symbol is flashing, the stimulus is significantly reduced.
symbol is flashing, the stimulus is significantly reduced.
Beauty Series Accessories:
You can purchase all accessories at www.bmrbeautyseries.com. Under no circumstances
should anything other than
accessories be used with your
BMR LIFT
390) or charger. Any others may not be compatible with the product and could degrade
the minimum safety levels.
Garment (E:70)
BMR TUMMY LIFT
GelPads:
•1 large GelPad (Type 706 or 709)
BMR TUMMY LIFT
•2 small GelPads (Type 707 or 710)
Garment (E:30)
BMR BOTTOM LIFT
GelPads:
•4 large GelPads (Type 716)
BMR BOTTOM LIFT
Garments (E:60)
BMR ARM LIFT
GelPads:
•4 small GelPads (Type 715)
BMR ARM LIFT
Belt Extender
Battery Pack (3.6V, NiMH)
Battery Charger (EU - 2504-0303)
Battery Charger (US/Japan - 2504-0302)
Charger complies to IEC 60950 and UL 1950, Input: nominal voltage 100-240V, frequency
50-60Hz, current 90mA. Output: 6V / 550mA. Do not use any other chargers or any other
power supplies.
Product Type: 394
Intended use: Muscle stimulator
Classification: Internally powered equipment, Class II charger, Type BF applied parts.
Waveform: Symmetrical bi-phasic square waveform when measured into a resistive load.
Environmental Specifications:
Operating:
Temperature Range: 0 - 35°C (32 - 95°F)
Humidity: 20 - 65 % RH
Storage:
Temperature Range: 0 - 35°C (32 - 95°F)
Humidity: 20 - 85 % RH
Garment Materials:
Outer material: 80% Nylon, 20% polyester; Inner Material: 100% Polyester laminated to
100% polyurethane; Binding: 100% Nylon; Hook & Loop: 100% Nylon; Foam: EVA;
Elasticated Straps: 100% Nylon.
14
All manuals and user guides at all-guides.com
is not required for maintenance purposes.
BMR
.
BMR
. These can be purchased
ACTIVATOR
ACTIVATOR FOR LIFT
Description of your
There are a number of technical markings on your
The
and garments are manufactured for Bio-Medical Research Ltd, Parkmore
ACTIVATOR
Business Park West, Galway, Ireland.
The
requires 1 x 3.6 volt (NiMH) DC battery pack DC is indicated by the symbol:
ACTIVATOR
The Output Frequency indicates the number of pulses per second transmitted by the
This is measured in hertz, which is indicated by "Hz".
or your local
Output (RMSA) means the maximum output root mean square current for each channel.
Output (RMSV) means the maximum output root mean square voltage for each channel.
!
This symbol means "Attention, consult the accompanying documents".
This symbol means type BF applied parts.
This symbol on your
Medical Device Directive (93/ 42/EEC). 0366 is the number of the notified body (VDE).
0366
(Type
SN stands for 'serial number'. Inside the box, on the back of the
specific to this
where "P" denotes 2010, "Q" denotes 2011 etc.
The garments' batch number is represented on their packaging, by the number corresponding
with the LOT symbol.
Rated Outputs – Voltage/Currents: E:60
Parameter
Output RMSV
Output RMSA
Output Frequency
DC Component: Approx.
Positive Pulse Width:
Negative Pulse Width:
Interphase Interval:
Attention:
The
can deliver currents in excess of 2mA/cm
ACTIVATOR
At the end of the product lifecycle, do not throw this product into the normal household
garbage, but bring it to a collection point for the recycling of electronic equipment.
Some product materials can be re-used if you bring them to a recycling point. By re-using
some parts or raw materials from used products you make an important contribution
to the protection of the environment. Please contact your local authorities if you need
more information about collection points in your area.
Waste Electrical and Electronic Equipment can have potentially harmful effects on the
environment. Incorrect disposal can cause harmful toxins to build up in the air, water
and soil and can be harmful to human health.
's symbols:
ACTIVATOR
ACTIVATOR
is to indicate conformity to the requirements of the
ACTIVATOR
Conforms to UL STD 60601-1.
Certified to CAN/CSA.
STD C22.2 NO 601.1
. The letter preceding the number indicates the year of manufacture,
ACTIVATOR
500
1K
6.4V
12.7V
12.7mA
12.7mA
50Hz
50Hz
0 C
0 C
175-200µs
175-200µs
175-200µs
175-200µs
100 µs
100 µs
2
. These can be explained as follows:
.
ACTIVATOR
is the serial number
ACTIVATOR
1K5
12.7V
8.5mA
50-Hz
0 C
175-200µs
175-200µs
100 µs
if used with incorrect GelPads.