Always keep the catheter safety valve clean and the access tip on the connec-
ted tube sterile.
Take care to ensure that they do not come into contact with any un-sterile objects in order to avoid
contamination. The safety valve must be disinfected before each drainage.
While draining, make sure that the safety valve and the access tip are secure-
ly connected. Take care not to tug or pull on the connected tube.
Take particular care to ensure that the access tip is inserted in the safety valve in a sterile manner and
make sure that the connection is secure. It can lead to contamination if the parts become disconnected
by mistake. In case this occurs, disinfect the safety valve with a new alcohol wipe and dispose of the
disconnected drainage material. Use a new sterile drainova® Reservoir to continue with the drainage.
en
The drainova® clickFix can be used to prevent unintentional disconnection.
The drainova® clickFix is not contained in the drainova® Reservoir but can be ordered separately.
Connecting the drainova® Reservoir
1.
Remove the drainova® Silicone Cap or the drainova® Protective Cap:
Using the drainova® Silicone Cap:
Firmly hold the safety valve of the catheter and carefully remove the drainova® Silicone
Cap. Dispose of the cap.
Using the drainova® Protective Cap:
Firmly hold the safety valve of the catheter, rotate the drainova® Protective Cap anticlock-
wise and carefully remove. Dispose of the cap.
2.
Hold the safety valve in one hand and clean the area around the safety valve-opening with an
alcohol wipe or a suitable alcohol-based disinfectant.