Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Measurements in technical gases
If the "MEASUREMENT TECHN. GAS" mode is selected, the user can
only manually operate the instrument. In this case, the user will be
prompted to start the measurement with a special operation procedure
needed for measurements in technical gases. This mode is used to
consider the different viscosity of a technical gas compared to the
viscosity of ambient air. Following the menu in this mode, the
Dräger X-act 5000 will automatically be adjusted to the needed flow
parameter. The measurement result can be read out directly at the
tube. The tubes listed in Chap. 10 can be used for measurement.
Password-protected menu
The Dräger X-act 5000 offers a password-protected menu. The
following items can be selected:
calibrate flow ("CALIBRATE FLOW")
change of the SO
filter ("CHANGE SO3 FILTER")
3
The date for the filter change can be set. This date is for
information only.
select operating modes ("SELECT MEASURING MODE")
The operating modes can be activated or deactivated. At least one
operating mode must be activated. Default setting: All modes of
operation are activated.
Changing the sample time resolution ("SAMPLE TIME
RESOLUTION")
The default setting of the sample time resolution on delivery is in
steps of 15 minutes. This configuration can be changed to a
sample time resolution in steps of 1 minute.
select leak test interval ("MANDATORY LEAK TEST")
The duty to perform a leak test can be deactivated or can be
mandatory after a certain number of strokes. If it is mandatory, the
leak test will automatically prompt after the set number of strokes.
Default setting: Mandatory leak test after 1000 strokes (= 100 L).
Dräger X-act 5000
update tube data base ("UPDATE TUBE DATA BASE")
The tube database can be updated by scanning a special bar code
with the bar code reader of the instrument. To do so, follow the
information printed in the instructions for use of a new or modified
tube.
change password ("CHANGE PASSWORD")
Select language ("SELECT LANGUAGE")
Default setting: The display language is set to English.
2.3.4
Keys
The green "OK" key will be used to
switch on the instrument,
acknowledge / confirm a selection, setting, prompt of error
indication or the indication of end of measurement,
start / stop / continue a measurement / a leak test / skip of a leak
test,
activate the display backlight,
deactivate the sleep mode.
The blue arrow keys will be used to
select a menu / option, a measuring mode / option, offered
prompts,
set a number of strokes,
set the flow rate and time,
switch OFF the instrument,
exit from a menu / option,
activate the display backlight,
deactivate the sleep mode.
The arrow keys will be disabled when a measurement is performed.
Description
9

Publicité

Table des Matières
loading

Table des Matières