Text/Code
Cause
100
ROM CRC fault
101
RAM Integrity fault
102
Info flash fault
103
Charger communication
fault
104
Boot loader
communication fault
105
Motor does not start
106
Motor does not stop
107
Motor too slow
108
Motor too fast
109
APS sensor out of range Contact DrägerService.
110
DPS sensor out of range Contact DrägerService.
111
DPS sensor fault
112
Software integrity fault
113
Data integrity fault
114
Volume out of range
Dräger X-act 5000
Remedy
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Remove and reinstall
rechargeable battery /
battery pack or contact
DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Calibrate flow or contact
DrägerService.
1) The bar code reader can be checked by scanning the following
test bar code. When the bar code reader works properly, the
display will indicate "1234567 TEST OK".
Test bar code:
Errors and warnings