en
Installation
Electrical Connection
•
Mains operation: Insert the input plug (h) of the mains adaptor (g) into the
electrical connection socket (i) of the unit (Ill. 1). The mains adaptor must
correspond to the local mains electricity supply. Important: Remove the
batteries before plugging the unit into the mains.
•
Battery operation: Make sure the mains adaptor is disconnected from the mains
and from the unit. Release the latch (Ill. 2)and slide open the battery cover. Re-
quired: 6 LR 06 / AA batteries. Check the + / - mark orientation of the batteries
(Ill. 3) and insert the batteries. Close the battery compartment cover and secure
by re-engaging the latch. The average working life when adjusting the vacuum
and cycle is ca. 5 hours (Duracell batteries).
Connecting the PumpSet
Use of the PumpSet is detailed in the separate User Manual: "ARDO PumpSet"
•
The PumpSet is an applied part of Calypso.
•
To use the single PumpSet: Attach the tube connector (c) to the unit (Ill 4).
The second opening of the tube connector remains closed.
•
To use the double PumpSet: Attach the tube adaptor (c) (with both
PumpSets attached) to the unit (Ill 5).
Operation
Vacuum and cycle can be adjusted independently of each other. Select the adjust-
ment which is the most comfortable for you.
Switch on / switch off
Use of the "ON / OFF switch (Ill. 6)
At switch-on, Calypso pumps at the lowest
intensity of vacuum (50mbar) and the lowest
cycle level (30 cycles per minute).
Operational Display: The operational display
has a green backlit LED. The display turns off
when the unit is turned off. The guarantee
is time-display related. This means that the
display will turn from green to orange after
400 hours of use.
The On and Off works in the same way for
both mains- and battery- powered units.
Regulation of the vacuum;
display (Ill. 7)
+
6
-
+
-
d
Touch the up-arrow to
intensify the vacuum. Touch
the ▼down-arrow to reduce
the vacuum.
Each touch changes the inten-
sity of the vacuum by one level.
The range of possible adjust-
ment lies between 50mbar
and 330 mbar (5- 33 kPa.).
(See Section 4: "Useful Tips ").
13
Cycle regulation; display
(Ill. 8)
7
Touch the ▼ (up-arrow) to
step-up the cycles one or
more level(s). Touch the ▼
(down-arrow) to reduce the
cycle one or more level(s).
Each touch changes the cycle
level by one step.
The range of possible adjust-
ment lies between 30 and 60
cycles per minute.
(See Section 4: "Useful Tips ").
Ardo medical AG
en
1
2
3
4
5
+
8
-
+
-