Accompanying Documents
The appliance is complied with the following emission and immunity standard, and test
level:
Immunity test
Electrosta�c
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interrup�ons and
voltage varia�ons
on power supply i
nput lines
IEC 61000-4-11
Power frequency
magne�c field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE UT is the a.c. mians voltage prior to applica�on of the test level.
03211_de-en-fr-nl_GM_A5_V4.indb 66
03211_de-en-fr-nl_GM_A5_V4.indb 66
IEC 60601 test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repe��on frequency
±0.5 kV, ±1 kV differen�al mode
±0.5 kV, ±1 kV, ±2 kV common
mode
0 % UT; 0,5 cycle. At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
Not Applicable
100 kHz repe��on frequency
±0.5 kV, ±1 kV differen�al mode
Not Applicable
0 % UT; 0,5 cycle. At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Radia
IEC61
(Test s
for EN
PORT
to RF
comm
equip