E
Error
Possible cause
The motion control
detected movement
during the measure-
ment.
Measured
Too little rest before
values/pulse
taking the measure-
values are
ment. Movement or
extremely high
talking during the
or low.
measurement.
Incorrect measure-
ment posture
Measurements
Blood pressure is not
taken in direct
a fixed value. It can go
succession
up or down by more
produce diffe-
than 20 mmHg within
rent measured
a few minutes.
values.
Measurement
Device detects a fault
was interrupted,
or that the diastolic
the cuff deflated
value is too low.
to 0 and the
Movement during
measurement
measurement
restarted.
No display after
Batteries inserted
device is swit-
incorrectly
ched on
Batteries flat
Battery contacts dirty.
Dashes in
No measured values
display after
have been stored.
pressing Memo-
ry button
Measurement is
Batteries flat
interrupted
Batteries flat
The time is
visible in the
display, but the
measurement
is interrupted
and the symbol
appears.
EN-58
Technical information
Remedy
- Repeat the measurement
- Do not move your arm
- Do not speak
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 43.
Repeat the measurement and
maintain the correct posture.
(Page 47)
In order to obtain compara-
ble results, always carry out
the measurements under the
same conditions.
Device shows no fault. Repeat
measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes' rest.
Check position of batteries.
Replace batteries.
Clean battery contacts with a
dry cloth.
Repeat measurement
Replace batteries.
Replace batteries.
Technical information
Error
Possible cause
Date and time
Function was uninten-
are not visible in
tionally deactivated
the display
2. Customer service
Device repairs may only be performed by the manufacturer or an
expressly authorised body. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax.: +49 (0) 9342/924080
Email: info@uebe.com
Website: www.uebe.com
3. Technical data
Model:
visomat
®
Size:
L = 127.3 mm x W = 162.3 mm x H = 96 mm
Weight:
approx. 338 g without batteries and cuff
Display:
LCD display (liquid crystal display) for mea-
sured values and check displays
Memory:
2 memories each capable of storing 120
readings
Measurement method: Oscillometric determination of systole, diastole
and pulse
Clinical trial:
EN 1060-4:2004 September 2004,
Minutes of the Deutsche Hochdruckliga Ger-
man Hypertension League December 2005,
Minutes of the European Society of Hyperten-
sion July 2008,
ISO 81060-2:2009 September 2014
Reference method of
Auscultatory measurement
the clinical test:
Remedy
Activate the date and time
function as described on
page 45.
comfort 20/40
EN-59
E