E
Cuff pressure range:
3-300 mmHg
Measurement range:
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse measurement: 40-160 pulses/minute
Measurement preci-
Blood pressure measurement (cuff pressure):
sion:
± 3 mmHg
Pulse rate: ± 5%
Power supply:
4 x 1.5 V AA LR6 alkaline manganese batteries
Optional: visomat
DC, 500 mA
Power consumption:
max. 4 W
Cuff:
Cuff type 2040 for an arm circumference of
23-43 cm, applied part type BF
Operating conditions:
Ambient temperature 10°C to 40°C, relative
humidity 15% to 85% non-condensing, air
pressure 700 to 1060 hPa
Conditions for storage
Ambient temperature -20°C to +60°C
and transport:
Relative humidity 10% to 95%, non-condensing
Inflation and deflation: Regulated automatically
IP rating:
IP 20: Protected against solid foreign bodies
with a diameter from 12.5 mm, no protection
against water.
Protection against
Internal power supply, applied part type BF
electric shock:
(cuff)
Operating mode:
Continuous operation
Classification:
Internal power supply using batteries.

4. Original spare parts and accessories

The following original spare parts or accessories are available
through specialist retailers:
Universal cuff 23-43 cm, type 2040
Part no. 2404001
PZN 01021375
EN-60
Technical information
mains adapter, output 6 V
®
Technical information
visomat
mains adapter
®
Part no. 2400020
PZN 11870684

5. Metrological inspection

Generally, it is recommended that a metrological inspection is
performed every two years. However, professional users in Ger-
many are obligated to do so in accordance with the "Medical
Products Operator Ordinance" (Medizinprodukte-Betreiberver-
ordnung).
This can be performed either by UEBE Medical GmbH, an author-
ity responsible for metrology or an authorised maintenance ser-
vice. Please refer to your national regulations.
Upon request, responsible authorities or authorised mainte-
nance services receive a "Test instruction for metrological inspec-
tion" from the manufacturer.
Please only submit the device for metrological inspection togeth-
er with the cuff and the instructions for use.
Important: No modifications, e.g. opening the device (except to
replace the batteries), may be made to this device without the
manufacturer's permission.

6. Explanation of symbols

This product complies with Council Directive 93/42/EEC of
5 September 2007 concerning medical devices and bears
0123
the mark CE 0123 (TÜV SÜD Product Service GmbH).
Degree of protection against electric shock TYPE BF
Subject to technical modifications.