Télécharger Imprimer la page

Citizen CHU304 Manuel D'instructions page 17

Publicité

SPECIFICATIONS
Model Number:
Measurement system:
Display
Measurement Localization
Cuff:
Cuff circumference range:
Cuff pressure:
Measurement range:
Accuracy:
LCD displays:
Switch:
Inflation:
Deflation:
Rated voltage:
Exhaust:
Power supply:
Battery duration:
Automatic Power Off function:
Main Unit Dimensions:
Cuff Size:
Weight:
Operating conditions:
Storage conditions:
Electric shock protection:
Degree of Protection
Mode of operation:
Memory:
Service Life:
Cuff's Life:
Protection against ingress of water:
Accessories:
Optional accessaries:
* Applied part for this device is Cuff.
* A range in barometric pressure 105 kPa to 80 kPa
* A range in altitude -300 m to 2000 m
* The Electronic Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN 60601-
1:2006/AC:2010 (Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance), IEC 60601-1-2: 2007/EN 60601-1-2:2007/AC:2010 (Medical electrical equipment – Part
1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic
compatibility - Requirements and tests), EN 1060 -1: 1995 + A1:2002 + A2: 2009 (Non-invasive
sphygmomanometers - Part 1: General requirements), EN 1060-3:1997 + A1: 2005 + A2: 2009 (Non-
invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure
measuring systems). IEC 80601–2–30: 2009+Cor. 2010 (Medical electrical equipment – Part 2–30: Particular
requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers)
00̲CHU304̲All.indb Eng-16
00̲CHU304̲All.indb Eng-16
CHU304, CHU305
Oscillometric method
Digital display type
Upper Arm
Soft cuff
22 cm – 30 cm (Cuff), 22 cm – 42 cm (Wide Range Cuff)
0 – 300 mmHg
Systolic: 60 – 260 mmHg
Pressure
Diastolic: 40 – 199 mmHg
Pulse
40 to 180 pulse/min
Pressure
±3 mmHg
Pulse
±5% of reading
Pressure
3 digits
Pulse
3 digits
: Measurement icon
: Ready to inflate indicator
: Battery icon
Icons
: Irregular Heartbeat (IHB) Indicator
: Memory icon
: Hypertension Classification Indicator
2 (START/STOP, MEMORY)
Automatic Inflation by internal pump
Automatic speed deflation system.
6V DC
Electromagnetic quick exhaust valve
4 X 15V
Approx. 800 times (Alkaline)
Approx. 1 min. (after activated)
138 (W) x 54 (H) x 95 (D) mm
Approx. 148 × 510 mm
Cuff
Wide Range Cuff
Approx. 163 × 580 mm
Unit: Approx. 207g w/o batteries
Cuff: Approx. 128g, Wide Range Cuff: Approx. 134g
Temperature
10°C – 40°C
90%RH
Humidity
Temperature
-20°C – 55°C
Humidity
90%RH
Internal power unit
Type BF applied part
Continuous operation
1 × 99 readings, Average of last 3 readings
10,000 measurements or 3 years
Approx. 1000 times
IPX0
Set includes an cuff, 4 AA batteries (LR6) for the
monitor, Instruction Manual
Cuff 22 cm – 30 cm, Wide Range Cuff 22 cm – 42 cm
- Eng 16 -
(
: direct current)
SIZE AA batteries (LR6)