• Place the medicine container (2) in a
position that allows you to apply a few
drops of ethyl alcohol to the mesh (3).
Allow it to work for 10 minutes.
• Finally, clean all the parts again under
running water (except for the mesh
atomiser).
Important
Ensure that you do not touch the
mesh (3), as this may destroy it. The
mesh must not be rinsed under run-
ning water, but in a vessel filled with
clean water. Otherwise the mesh
may become damaged.
• If any traces of medical solutions or im-
purities remain on the gold-coloured
contacts on the device and nebulizer,
use cotton buds dampened with ethyl
alcohol to remove them.
• Clean the outside of the mesh and out-
side of the medicine container with a
cotton bud dampened with ethyl alco-
hol.
Drying
• Place the individual components on a
dry, clean and absorbent surface and
allow to dry completely (for at least 4
hours).
All manuals and user guides at all-guides.com
31
Note:
Ensure that all parts are thoroughly dried af-
ter cleaning, otherwise there is an increased
risk of bacterial growth.
Once all components are fully dry, reas-
semble the nebulizer and place the com-
ponents in a dry, sealed container.
Material resistance
• When choosing a cleaning or disinfect-
ing agent, note the following:
Only use mild cleaning or disinfectant
agents in the quantities recommended
by the manufacturer.
• As with all plastic components, frequent
use and cleaning of the accessories can
lead to a certain amount of wear. Over
time, this can alter the aerosol proper-
ties and may eventually affect the effi-
ciency of therapy. We therefore recom-
mend that you replace the nebulizer
and accessories at least once a year.
Storage
• Do not store the nebulizer in a damp
atmosphere (e.g. in the bathroom) or
transport it together with damp objects.
• Store and transport the nebulizer away
from direct sunlight.