Important
Before you clean the device, always
switch it off, unplug it and let it cool off.
Do not use abrasive detergents and never
immerse the device in water.
Important
• Ensure that no water penetrates inside
the device!
• Do not put the device or accessories in a
dishwasher.
• Do not touch the device with wet hands
while it is plugged in; do not allow any
water to be sprayed onto the device.
The device must be operated only
when it is completely dry.
• Do not spray any liquids into the venti-
lation slots.
Any liquid that penetrates into the de-
vice can damage the electrical parts or
other nebulizer components and impair
the function of the device.
Disinfection
• First clean the nebulizer and the acces-
sories as described in the Cleaning sec-
tion. You can then continue with disin-
fection.
• Pour 8 ml ethyl alcohol into the medi-
cine container (2). Close the container.
Leave the alcohol in the container for at
least 10 minutes.
• For better disinfection, shake the device
gently from time to time.
• Afterwards, pour the ethyl alcohol out of
the container.
• Repeat this process, using water this
time.
All manuals and user guides at all-guides.com
30
Disassembly
• To continue with cleaning its individual
components, disassemble the nebulizer
(2, 3):
• Remove the medicine container (2) and
mesh nebulizer (3) by pressing both
brackets inwards and drawing the de-
vice out.
• Remove the mesh nebulizer (3) from
the medicine container (2) by turning it
anticlockwise and drawing it out.
• The nebulizer is later reassembled in
the reverse order.