10.2.4.5.
Lecture du seuil d'alarme CMD-5
Description : Lecture de la valeur du seuil d'alarme
Commande du lien descendant :
Paramètre
-
topic : /prodcutID/deviceID/function/invoke
{
"messageId":"1574326733176995841",
"deviceId":"173072083110001",
"timestamp":1664183717422,
"functionId": "CMD5",
"messageType": "INVOKE_FUNCTION",
"inputs":[{"cmd":"CMD-5"}]
}
Uplink ACK (message d'accusé de réception de la commande qui informe le serveur que la
commande a été reçue) :
topic:/prodcutID/deviceID/function/invoke/reply
{
"messageId":"1574326733176995841",
"cmd": "CMD-999",
"output": "success"
}
Données du lien ascendant :
Paramètre
alarm_threshold_0.3um
alarm_threshold_0.5um
alarm_threshold_1.0um
alarm_threshold_5.0um
alarm_threshold_10um
Unit
topic : /prodcutID/deviceID/properties/report
{
"cmd": "CMD-105",
"desired":{
"alarm_threshold_0,3um": 123,
"alarm_threshold _0,5um" : 123,
"alarm_threshold_1.0um" : 123,
"alarm_threshold_5.0um" : 123,
"alarm_threshold_10um" : 123,
"Unit": "pcs/m3"
}
}
Type
-
Type
Number
Number
Number
Number
Number
String
PCE Instruments
25
Description
-
Description
Valeur seuil des particules >0,3 μm
Valeur seuil des particules >0,3 μm
Valeur seuil des particules >0,3 μm
Valeur seuil des particules >0,3 μm
Valeur seuil des particules >0,3 μm
3