MABTECH INSTRUMENTS
germicide. Hydrogen peroxide can be corrosive and affect the skin and mucous
membrane, and precautions must be taken when handling the chemical.
Special instructions Hazard Group 3 or 4
1. Instruments used in Biosafety/Containment Level 3 or 4 laboratories must be
decontaminated by the Customer using an internationally approved sterilization
procedure.
2. The Customer must then move the instrument to a laboratory with Containment Level 1
or 2. Mabtech personnel are not allowed to enter into a Containment Level 3 or 4
laboratory to perform maintenance or service.
3. It may not be possible for Mabtech to service or transport instruments used in Level 3 or
4 laboratories.
Radioactive Materials
1. Apply an industry standard radioactive decontaminant (Radiacwash® or similar) to the
instrument and wipe surfaces as directed by the manufacturer's instructions.
2. Survey the instrument with an appropriate radioactive-measuring instrument.
3. Satisfactory decontamination is defined as survey results at or below background level,
or in the US only, for service work excluding transportation, level designated to be clean
or safe in the Customer's regulatory approved Site Radioactive materials License.
Hazardous Chemicals
1. Areas exposed to hazardous chemicals should be washed with an acceptable solvent
such as ethanol or isopropanol.
2. Rinse the surface with detergent and water.
Please note that Mabtech cannot accept any instrument or parts that may be
contaminated with viable biological agents, harmful quantities of chemicals or
radioactive materials.
4