MD300W512
● methylene blue
● Indigo carmine
(3) Extremely high illumination could affect the SpO
sensor.
(4) Other factors
a) High-frequency electrosurgical interference from external units, including defibrillators.
b) Placement of a sensor on an extremity that currently has been placed a blood pressure cuff, arterial catheter, or intravascular
line;
c) The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia;
d) An arterial occlusion proximal to the sensor.
Warnings!
Do not use the SpO
2
Excessive patient movement may cause inaccurate measurements.
Tissue damage can be caused by incorrect operation or misusing sensor; for example, by wrapping the sensor too
tight. Inspect the sensor site to ensure the skin's integrity and the adhesion position of the sensor is correct. More
frequent inspection should be taken if necessary.
Loss of pulse signal can occur in any of the following situations:
a) The sensor is too tight;
b) There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or sunlight;
c) Do not use the device on the same arm when taking a blood pressure reading.
After measurement, please take off your finger and press and hold the power button to turn off the device.
All manuals and user guides at all-guides.com
sensor with exposed optical components.
measurement. Use a semi-translucent or opaque cover to shield the