Télécharger Imprimer la page

Ambu Disposable Face Mask Mode D'emploi page 4

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34
1.6. Warnings and cautions
Failure to observe these precautions may result in inefficient ventilation of the patient,
cross-infection or damage to the equipment.
WARNINGS
1. For single patient multiple use only. Use on other patients can cause cross infection.
2. Only to be used by intended users who are familiar with the content of this manual,
as incorrect use may harm the patient.
3. Always ensure to remove all packaging material from the Ambu Disposable Face Mask
prior to use as the pouch could block the patient's airways and prevent ventilation.
4. Always visually inspect the product after unpacking, assembly and prior to use as defects
and foreign matter can lead to no or reduced ventilation of the patient.
5. Do not use the product if inspection fails, as this can lead to no or reduced ventilation.
6. Do not use the product if contaminated by external sources, as this can cause infection.
7. Always inspect the proper inflation of the Ambu Disposable Face Mask cushion before
use, as poor sealing might lead to insufficient ventilation of the patient.
8. Always ensure correct positioning and sealing of the Ambu Disposable Face Mask,
as improper sealing can lead to spreading of airborne infectious disease to the user.
9. Do not use the Ambu Disposable Face Mask for more than 4 accumulated hours over
a maximum time span of 1 week, in order to avoid the risk of infection.
10. Do not reuse the Ambu Disposable Face Mask if visible residues are left inside the device
in order to avoid the risk of infection or malfunction.
11. Always ensure correct positioning of the Ambu Disposable Face Mask on the patient
by performing, with either hand, an appropriate jaw thrust through relevant hand
grip strategy, or by head straps connected to the hook rings. Reposition if applicable,
as improper sealing of the face mask could lead to insufficient ventilation of the patient.
12. Do not apply excessive force on the face mask when keeping it in position, as it might
lead to pressure marks on the patient's face.
13. Always visually inspect the inflation and sealing properties of the disposable face mask
cushion while in use on the patient, as improper sealing might lead to leakage and
reduced or no ventilation of the patient.
14. When using supplemental oxygen, do not allow smoking or use the device near an open
flame, oil, grease, other flammable chemicals or equipment and tools, which cause sparks,
due to the risk of fire and/or explosion.
15. Professionals performing the procedure should assess the choice of face mask size
and compatible devices (e.g. resuscitator, etc.) in accordance with the patient's specific
condition(s), as incorrect use may harm the patient.
16. Do not use the product when ventilating patients with severe facial trauma and/or eye
injury due to the risk of improper sealing and aggravation of injury unless a medical
assessment indicates the necessity. Switch to an alternative method for directing air to
the patient, if available.