17
Technical Information
Interconnections
Connection J11 – Purge Flow Meter Connector
Figure 15
Tab. 5
PIN
1-3
5-6
WARNING!
Use this connector (J11) only with the Agilent solid state flowmeter model
969-9114 or 969-9115. Any other connection could lead to serious injury.
220/268
J11 connectors
DESCRIPTION
Flow meter power supply: 12 Vdc, 10 mA
(pin 1 negative - pin 3 positive)
Analog signal with a voltage value proportional to the
flow.
Between 0 and 3V it is linearly proportional to between 0
and 100 sccm (pin 6 negative, pin 5 positive)
Turbo-V 1000 ICE Controller User Manual / 87-900-905-01 (M)