Télécharger Imprimer la page

Tecno-gaz ANDROMEDA VACUUM xp Mode D'emploi page 36

Masquer les pouces Voir aussi pour ANDROMEDA VACUUM xp:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
14
The autoclave is programmed by the factory to stop heating after 60 inactivity minutes
(STAND BY). So it is possible to carry out sterilisation cycles even in the absence of the
operator, because once the cycle has been ended after 60 minutes, the autoclave is
automatically switched off. All LEDS turn off and only the main ON/OFF switch remains on.
To check that sterilisation has taken place, the operator must press the "START/STOP"
button to reset the autoclave. In this case, the unit will display the last sterilisation cycle run
by the operator.
In case the "TIME/MESS." display shows "END", this means that sterilisation was
successful. If an alarm or an error message are displayed, see PAR.16 for more help about
possible troubles occurred during sterilisation.
15
CAUTION:
Disconnect always the main supply before each technical intervention
15.1 Daily Routine Maintenance
CLEANING THE DOOR SEAL
Clean the seal on the door and on the outer edge of the sterilisation chamber sealing the
gasket using a wet cloth or the soft part of the supplied sponge.
This cleaning operation must be carried out to remove possible dirt which can cause a
pressure loss in the sterilisation chamber and a possible seal cut.
Check the water level in the tank(see PAR.9)
15.2 Weekly Routine Maintenance
CLEANING THE STERILIZATION CHAMBER
To remove the deposits on the chamber floor, clean with the abrasive part of the
supplied sponge after removing the tray-holder.
To moisten the sponge, only use distilled or demineralized water.
Empty the water tank (see PAR.11).
14
NIGHT CYCLE
MAINTENANCE

Publicité

loading

Ce manuel est également adapté pour:

2053-s2053sippc8