14. When the autoclave is embedded it is necessary to follow these instructions:
Place the rilsan pipe (Diam.8) supplied with the autoclave in POS.1 Photo A or in the
fitting pipe A (if present)
Discharge by gravity the other extremity of the pipe in an open container, to allow the
discharge of the steam. THE PIPE MUST NEVER GET IN TOUCH WITH THE WATER
INSIDE THE CONDENSE CONTAINER.
15. Make sure that the electrical system to which the device is connected is produced
according to the laws in force and has the suitable dimension for the device features;
see PAR.2 of this manual.
16. Plug
the
unit
MONOPHASE+EARTH,frequency 50/60Hz, current 16A, as shown on the identification
plate place in the rear side of the device.
17. Never connect the device pin to reductions of any type.
18. Load the filling tank as explained on PAR.9, of this manual.
19. After the installation, carry out a LOADED sterilisation cycle as explained on PAR.12 of
this manual.
LEGEND
1 – Drain Exceeding clean water – condensate discharge
2 – Discharge of clean water
3 – Safety Valve
A – Condensate Discharge (IF PRESENT)
8
to
a
socket
REAR PANNEL CONNECTIONS
PHOTO A
with
a
rated
voltage
of
230V 10%