11. During the pre-heating phase when the autoclave is heating to reach the right pressure,
on the "TIME/MESS." display the letter H is displayed followed by the blinking time. After
reaching the selected sterilisation parameters, on the "TIME/MESS." display, the letter H
disappears and a blinking letter S is followed by the sterilisation time which
progressively decreases up to the end of the cycle.
12. After the sterilisation time, the "TIME/MESS." display, visualises letter A followed by the
drying time which indicates when the autoclave is in drying phase and then the vacuum
pump starts working. The vacuum pump is started to suck the rest steam enabling a
better drying.
13. In case after starting the sterilisation cycle you wish to stop it, proceed as follows: Press
the "START/STOP" button, wait that the pressure indicated by the "PRESSURE"
display, is at 0 bar. On the "TIME/MESS." display, the writing "Int" (interrupted cycle) is
displayed, press the "13A" button, to cancel the alarm. Press the "14A" button, to
activate the door unlocking device: now it is possible to open the door with the handle.
14. If at the end of the sterilisation cycle on the "TIME/MESS." display, an alarm code is
displayed, for instance: (AL6, AL7, etc.... see PAR. 16), this means that the sterilisation
cycle has not taken place in the right way, so it is necessary to repeat the cycle.
15. At the end of the cycle on the "TIME/MESS." display the writing "END" is displayed
followed by an acoustic sign: Now it is possible to open the door by following the
warnings below.
16. Before opening the door, make sure that there is no pressure inside the sterilisation
chamber; the "PRESSURE" display, must indicate 0 bar. If the pressure value is higher
than 0, is not possible opening the door.
17. Use protective gloves to remove the load from the sterilization chamber.
18. When opening the door, do not stay above or in front of it to avoid burn due to the
vapour coming outs.
19. When opening the door it is normal to find residual water inside the door seal.
20. If the sterilisation chamber is not opened after the end of the sterilisation cycle, the inner
vapour tends to condense on the closing door, thus forming condensate drops which fall
on the floor while opening.
21. Remove the instruments from the autoclave.