MEASure:AMPLitude?
MEASure:AC?
MEASure:VOLT[:DC]?
MEASure:RISE:OVERshoot?
MEASure:FALL:OVERshoot?
MEASure:RISE:TIME?
or
MEASure:RTIME?
MEASure:FALL:TIME?
or
MEASure:FTIME?
MEASure:PWIDth?
MEASure:NWIDth?
(Query)
To the question MEAS: AMPLitude? <INT{1|2|3}> the instrument returns the amplitude of the selected
signal.
Response format:
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:AC? <INT{1|2|3}> the instrument returns the RMS voltage of the selected signal.
Response format:
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:VOLT? <INT{1|2|3}> the instrument returns the average value of the selected
signal.
Response format:
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:RISE:OVER? <INT{1|2|3}> the instrument returns the positive overshoot of the
selected signal.
Response format:
value in format <NR2> expressed in percent.
(Query)
To the question MEAS:FALL:OVER? <INT{1|2|3}> the instrument returns the negative overshoot of
the selected signal.
Response format:
value in format <NR2> expressed in percent.
(Query)
To the question MEAS:RISE:TIME? <INT{1|2|3}> the instrument returns the rise time of the selected
signal.
Response format:
value in format <NR3> expressed in second.
(Query)
To the question MEAS:FALL:TIME? <INT{1|2|3}> the instrument returns the fall time of the selected
signal.
Response format:
value in format <NR3> expressed in second.
(Query)
To the question MEAS:PWID? <INT{1|2|3}> the instrument returns the positive pulse width of the
selected signal.
Response format:
value in format <NR3> expressed in second.
(Query)
To the question MEAS:NWID? <INT{1|2|3}> the instrument returns the negatitive pulse width of the
selected signal.
Response format:
value in format <NR3> expressed in second.
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
85

Publicité

loading

Ce manuel est également adapté pour:

Handscope ca 922 - 20 mhz