3.2. Remote control commands
148
3.2.2.25. Mode.Parameter.Evaluation.EPC
Mode.Parameter.Evaluation.Recognition.Selected
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:
etc. up to 9 fix EP's
Mode.Parameter.Evaluation.pK
Parameters for DET and MET: Evaluation of the EP's, see page 21.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or uA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 23.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.26. Mode.Parameter.Presel.IReq
Mode.Parameter.Presel.SReq
Parameters for DET, MET, SET, KFT, STAT, DOS, DOC, MEAS: Automatic in-
quiry after the start of the determination. From such an inquiry, the determina-
tion continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.122) or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.27. Mode.Parameter.Presel.LimSmplSize.Status
Mode.Parameter.Presel.LimSmplSize.LoLim
Mode.Parameter.Presel.LimSmplSize.UpLim
Limit control for the sample size.
DET: 0...5...200
MET pH: 0.1...0.50...9.99
U, Ipol:
1...30...999
Upol:
0.1...2...99.9
all, greatest,
last, window, OFF
pH:
0...±20.00, OFF
U, Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Input range as LowLim
0...±20.00, OFF
U; Ipol: 0...±2000, OFF
Upol: 0...±200.0, OFF
ON, OFF
id1, id1&2, all, OFF
value, unit, all, OFF
ON, OFF
0.0...999 999
0.0...999 999
799 GPT Titrino