■ Plug the Exelia Maintenance Adaptor into an outlet on a different circuit from the one to
which the disruptive equipment is connected.
■ In any case, whatever the context, you should conduct interoperability testing in a real
situation to find the correct setup and location.
WARNING
The Exelia Maintenance Adaptor with Exelia infusion modules should not be
used next to, or stacked with, other equipment as this could result in improper
operation. If adjacent use is necessary, observe the Exelia Maintenance Adaptor
with Exelia infusion modules and the other equipment to verify that they are
operating normally.
The following configuration has been tested: one Exelia Maintenance Adaptor
with one Exelia SP.
1.9.1 Electromagnetic emissions
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations
Flicker emissions IEC
61000-3-3
1.10 Notification of serious incidents
Any serious incident that has occurred in relation to the device should be reported to the
manufacturer and the competent authority.
Information and contact information:
Fresenius Kabi AG
Else-Kröner-Str. 1
61352 Bad Homburg - GERMANY
Tel: +49 (0) 6172 / 686-0
www.fresenius-kabi.com
1.11 Release notes
Date
February 2020
Compliance obtained
by the device
Group 1
Class B
Complies
Class B
Complies
Creation
Electromagnetic environment - guidance
The Exelia equipment only uses RF energy for its
internal operation. Its RF emissions are therefore
very low and are not likely to cause any
interference in nearby electronic equipment.
The Exelia equipment is suitable for use in all
establishments, including hospital establishments
and those directly connected to the public
low-voltage power supply network.
Description
13