Output terminal -
connector
Symbol descriptions
WARNING
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of the device can lead to degraded performance,
safety hazards such as electric shock, increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and result in improper
operation.
1.8.1 Performance
The Exelia Maintenance Adaptor has no essential performance of its own, as defined in the
IEC 60601-1 standard.
1.9 Guidance and manufacturer's declaration on EMC
The Exelia Maintenance Adaptor has been tested in accordance with the electromagnetic
compatibility standards applicable to medical devices. Limitation of the emitted radiation
avoids undesirable interference with other equipment.
If the Exelia Maintenance Adaptor causes harmful interference, or if it is disrupted by external
interference, try the following:
■ Redirect or relocate the devices or disruptive equipment.
■ Change the routing of the cables.
■ Separate the power cords from the communication cables/signals.
■ Connect the Exelia Maintenance Adaptor to a protected/backed-up/filtered supply or
directly to the UPS circuit (Uninterruptible Power Supply).
■ Be careful with loops formed by communication cables and/or power circuits.
■ Use isolation transformers with respect to the hospital grid or insulated bridges.
■ Increase the distance between the Exelia Maintenance Adaptor and the disruptive
equipment.
15 V
(direct current) / 2 A
CE mark
Protection against electric shocks: class II
Warning (Refer to the Instructions For Use)
Refer to the Instructions For Use
In-door use only
Part included in a recycling process
European norms electrical certification
UL certification for US and Canada