Loading the appliance
Loading vaccines
When the temperature in the vaccine
compartment has stabilized, i.e. and the
temperature is between +2° and +8°C and
the compressor stops and starts, vaccines
can be loaded. The vaccines should be
placed and arranged in the baskets, and
remember to place the water filled ballast
lids at the top of the upper baskets. (fig. 5)
To ensure air-circulation and prevent too
low vaccine temperature, direct contact
to the inside walls must be avoided. The
Vaccine load in the bottom basket shall not
be higher than 250 mm.
Do not load lower baskets
above this height (250mm)