Filter Replacement - Drive DeVilbiss Healthcare AirForce Mini Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 123
EN
it to the air outlet (Pic. G) and switch on the device for 10-15
minutes.
!
Do not boil or autoclave the air hose the masks.
NEBULIZER REPLACEMENT
The nebulizer must be replaced after a long period of inactivity,
in case it shows deformations or breakings, or when the nebu-
lizer nozzle is obstructed by dry medicine, dust, etc.
DeVilbiss
recommends to replace the nebulizer after a period
between 6 months and 1 year, depending on the usage.
Use original nebulizers only.
!

FILTER REPLACEMENT

In normal usage conditions, the air filter must be re-
placed approximately after
a year.
D
air filter (10 – 12 treatments) and, if the filter shows a grey or
brown colour or is wet, replace it. Remove the filter (Pic. A-4)
and re- place it with a new one. Do not attempt to clean the
filter to reuse it.
Only use original filters.
!
Do not use the device without the filter.
26
eVilbiss
recommends to periodically check the
1
00 working hours or once

Publicité

Table des Matières
loading

Table des Matières