Centrifuge FC5513, FC5513R, FC5513L
2.3
Installation
Follow these steps:
•
Check whether the power supply corresponds with the one specified on the manufacturer's rating label,
which is located on the rear panel.
•
The line voltage circuit breaker is max. 10 A (type K) slow release for commonly used instruments.
•
In case of emergency, there must be an emergency switch off installed outside the room in order to
disconnect the power supply from the unit.
•
Connect the centrifuge to a grounded power receptacle.
•
Turn the instrument on, by using the mains power switch.
•
Open the lid by using the Stop / Lid button.
•
Remove the transport securing device of the motor.
2.4
Safety precautions during operation
•
Do not operate the centrifuge in case it is not installed correctly.
•
Do not lean on the centrifuge during operation.
•
Do not stay within the 30 cm clearance envelope longer than necessary for operational reasons.
•
Do not place any potentially hazardous materials within the 30 cm clearance envelope.
•
Do not operate the centrifuge when disassembled (e.g. without housing).
•
Do not run the centrifuge when mechanical or electrical components have been tampered with.
•
Do not use accessories such as rotors and buckets, which are not exclusively approved by OHAUS
Corporation, except commercially available centrifuge tubes made of glass or plastic.
•
Do not spin extremely corrosive substances, as they may damage or weaken the materials.
•
Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or mechanical damage.
The manufacturer is responsible for safety and reliability of the centrifuge, only if:
•
The unit is operated in accordance with this instruction manual.
•
Modifications, repairs, or other adjustments are performed by authorized personnel and the electrical
installation complies with the relevant electrical codes.