1. INTRODUCTION
This manual was created to help you install and learn all about the functional possibilities of the ASC scale.
The scales should be sited in a location that will not degrade the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The tables or floor must be rigid and not vibrate. Do not place near vibrating machinery.
Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.
Avoid high humidity that might cause condensation. Avoid direct contact with water. Do not spray or immerse the scales in
water.
Avoid air movement such as from fans or opening doors. Do not place near open windows.
Keep the scales clean.
Do not stack material on the scales when they are not in use.
2. MAIN TECHNICAL SPECIFICATIONS
Model
Plate
dimensions
ASC6
230 x 300
ASC15
230 x 300
ASC30
230 x 300
Stabilisation Time
Operative Temperature:
Power supply:
Battery life:
Recharge time:
Parameter Setup:
Display :
Case:
Zero range:
Auto zero at start-up:
3. INSTALLATION
a) Unpack then product.
b) If present (depending on the model), it is necessary to unscrew and remove the locking screw for the transportation
placed under the scale, as shown in the figure below:
Capacity
mm
kg
6
15
30
2 Seconds typical.
0°C - 40°C / 32°F - 104°F.
with fitted 240 Vac mains power adapter or with fitted battery
up to 70 hours of continuous use.
12 hours.
Calibration fully digital and programmable from the keyboard.
3 LCD backlit with 6 digits
ABS Plastic.
+/- 2 % of max load capacity.
+/- 10 % of max load capacity.
WARNINGS
Standard
Internal
Division
sensitivity
g
g
0,1
0,02
0,2
0,05
0,5
0,1
.
4
Min. APW
CE-M
recommended
Division
g
g
0,1
1/2
0,2
2/5
0,5
5/10