Instructions
Intended Use
This device is intended for medical pur-
poses to illuminate body surfaces. It is
used for non-invasive visual examination
of intact skin.
This battery-operated product is de-
signed for external examination only
within professional healthcare facilities
by medical professionals.
Check the correct operation of the de-
vice before use! Do not use the device if
there are visible signs of damage.
CAUTION: Do not look directly into the
LED light. Patients must close their eyes
during examinations.
In case of a serious incident with the use
of this device, notify DermLite imme-
diately and, if required by local regula-
tions, your national health authority.
CAUTION: Do not use the device in fire
or explosive risk area (e.g. oxygen-rich
environment).
WARNING: This product can expose
you to chemicals including methylene
chloride and hexavalent chromium,
which are known to the State of Cali-
fornia to cause cancer or reproductive
toxicity. For more information go to www.
P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC
Emissions and Immunity level require-
ments of the standard IEC 60601-1-
2:2014. The emission characteristics of
this equipment make it suitable for use in
professional healthcare environment as
well as residential environment (CISPR
11 Class B). This equipment offers ade-
quate protection to radio communication
service. In the rare event of interference
to the radio communication service, the
user might need to take mitigation mea-
sures, such as relocating or re-orienting
equipment.
WARNING: Use of this equipment adja-
cent to or stacked with other equipment
should be avoided because it could re-
sult in improper operation. If such use is
necessary, this equipment and the other
equipment should be observed to verify
that they are operating normally.
WARNING: Use of accessories other
than those provided by the manufac-
turer of this equipment could result in
increased electromagnetic emissions or
decreased electromagnetic immunity of
this equipment and result in improper
operation.
WARNING: Portable RF transmitters
should be used no closer than 30 cm (12
inches) to any part of the device. Other-
wise, degradation of the performance of
this equipment could result.
IMPORTANT: Before initial use, please
charge the internal lithium ion battery.
See "Battery and Charging" for details.
DermLite DL1 is a mobile device-com-
patible dermatoscope designed to view
skin lesions with high magnification
and clarity. Four LEDs provide bright,
even illumination, and a cross-polarized
spacer provides a glare-free image
by cancelling the reflected light from
the skin. A high quality, three-element
lens configuration, designed to provide
excellent results with leading mobile
devices, offers superb color correction
and reduced image distortion for images
rich in surface detail.
Your DL1 is equipped with a magnetic
adapter called MagnetiConnect® Clamp
("MCC"), designed for attaching DL1 to
most smartphones and tablets. To attach
MCC, position the clamp (MC) over the
main camera, then secure it by tighten-
ing the knob (KN). Attach the threaded
magnetic ring and place DL1 on it.
With its cross-polarized Spacer (B) and
glass contact plate (E) installed, place
DL1's contact plate directly on the skin.
Applying an immersion fluid onto the
skin may improve the image. Push the
power button (C) to activate the light;
push it again to turn it off. To conserve
power, your DL1 is designed to automat-
ically turn off after 2 minutes.
Glass Contact Plate
The unit's glass contact plate (E) may
be removed for cleaning or non-contact
imaging by rotating the bayonet ring (D)
counter-clockwise. Handle the glass
faceplate with great care.
Troubleshooting
Please check www.dermlite.com for the
most current troubleshooting informa-
tion. If your device requires servicing,
visit www.dermlite.com/service or con-
tact your local DermLite dealer.
Care and Maintenance
CAUTION: No modification of this equip-
ment is allowed.
Your device is designed for trouble-free
operation. Repairs shall be made only by
qualified service personnel.
Cleaning
The exterior of your device, except the
optical parts, may be wiped clean with
isopropyl alcohol (70% vol.) prior to
use on a patient. The lens should be
treated as high-quality photographic
equipment and should be cleaned with
standard lens cleaning equipment and
protected from harmful chemicals. Do
not use abrasive material on any part of
the equipment or immerse the device in
liquid. Do not autoclave.
Battery and Charging
CAUTION: This device uses a special
3.7V 180 mAh lithium ion battery, which
can only be purchased from DermLite or
from an authorized DermLite dealer. Do
not under any circumstances use anoth-
er battery other than the one designed
for this unit.
Prior to using the DL1, connect the wider
connector of the provided USB cable to
any IEC 60950-1 compliant (5V) USB
port (e.g. of a laptop, desktop computer,
or Apple charger) and the other end to
the DL1 Micro USB charging port (CP).
The charging indicator (G) will illuminate
red to indicate that the DL1 is being
charged. When fully charged, the indica-
tor turns green and the device will allow
up to an hour of continuous operation. If
less than 25% battery life remains, the
indicator blinks.
DermLite DL1 Basic components in-
clude:
DL1 dermatoscope (four white LEDs,
three-element lens, faceplate with 10mm
reticle, battery), cross-polarized spacer
with contact plate, charging cable, Mag-
netiConnect Clamp adapter
Warranty: 5 years for parts and labor.
The battery is warranted for 1 year.
Disposal
This device contains electronics and a
lithium-ion battery that must be separat-
ed for disposal and may not be disposed
of with general household waste. Please
observe local disposal regulations.