On-Site Test
VVX with analogue output 0.5...3.5 V
 Connect the supply voltage ( p. 23).
 Use a digital multimeter to check whether the supply voltage is present at pin 1 and pin 3.
 Check the voltage output (pin 2 and pin 3) with a digital multimeter:
 Without flow, the voltage output should be between 0.3...0.5 V.
 For a flow rate in the measuring range (see type plate), the voltage output should be
between 0.5...3.5 V depending on the flow rate.
4.5
4
3.5
3
2.5
2
1.5
1
0.5
0
6
Disposal
In accordance with Directives 2011/65/EU (RoHS) and 2012/19/EU (WEEE)*, the device
must be disposed of separately as electrical and electronic waste.
NO HOUSEHOLD WASTE
The device consists of various different materials. It must not be disposed of with
household waste.
 Take the device to your local recycling plant
or
 return the device to your supplier or to SIKA.