Guidelines and declaration of the manufacturer – electromagnetic immunity
EN
The model Medel Check is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Medel Check should
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
36
100042_MEDEL CHECK USER MANUAL_ ALL LANGUAGES.indd 36
All manuals and user guides at all-guides.com
ensure that it is used in such an environment.
IEC 60601
Compliance
Test level
level
3Vrms
Not applicable
150kHz to
80MHz
3V/m
3V/m
80MHz to
2,5GHz
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the model Medel
Check, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distances
d=1.2√P
d=1.2√P 80MHz to 800MHz
d=2.3 √P 800MHz to 2.5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b Interference
may occur in the vicinity of
equipment marked with the
following symbol:
16/05/2016 09:47:01