The model Medel Check is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Medel Check should
Immunity
test
Electrostatic
discharges
(ESD)
IEC 61000-4-2
Electrical fast
transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-
4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage before the application of the test level
100042_MEDEL CHECK USER MANUAL_ ALL LANGUAGES.indd 35
All manuals and user guides at all-guides.com
Guidance and manufacturer's declaration – electromagnetic immunity
ensure that it is used in such an environment.
IEC 60601 test level
± 6kV contact
± 8kV air
±2 kV for power
supply lines
.......
± 1kV in differential
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for
5 sec.
3 A/m
Compliance level
Electromagnetic
environment - guidance
± 6kV contact
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
± 8kV air
material, the relative humidi-
ty should be at least 30 %.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Not Applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the model
Medel Check requires conti-
nued operation during power
mains interruptions, it is
recommended that the model
Medel Check be powered
from an uninterruptible
power supply or a battery.
3 A/m
Power frequency magnetic
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or
hospital environment.
EN
35
16/05/2016 09:47:01