Troubleshooting
Symptom
Unit does not work
"Low pressure" alarm
Mattress does not
pump up
Patient bottoms out
No alternating pressu-
re takes place
Technical data
UNIT
Power supply
230 V, 50 Hz
protection class 1; IPX0;
type BF, AP/APG NO
Power consumption max. 12 Watt
Weight
2,2 kg
Dimensions
36.9 x 13.5 x 25 cm
Pressure range
30 - 60 mbar
Drive Medical GmbH & Co. KG | As per: 28.07.2023 | Errors and alterations excepted
Remedial action
• Check whether the plug is correctly inserted. Switch on the unit again.
• If the LED on the On/Off switch does not light up, check the voltage of the outlet or
insert the plug into another outlet.
• If the LED on the On/Off switch lights up but the unit does not work, please contact
your specialist dealer.
• Check whether the connection hose on the unit is correctly inserted.
• The CPR valve must be closed.
• Ensure that all cells are correctly connected to each other and that the connection
hoses are not twisted or kinked.
This indicates excessively low pressure in the air chambers. Increase the user weight on
the unit and perform a hand check to check whether the patient is sinking down.
• Check air filter (located on the back of the unit); please clean or replace it.
• If the unit continues to generate too little pressure, the unit must be replaced. Contact
your specialist dealer
• Ensure that the air supply hoses are not kinked or twisted.
• The static function must be switched off. When the static function is switched on, no
alternating pressure takes place.
MATTRESS REPLACEMENT SYSTEM
Dimensions 200 x 88 x 20 cm
without foam pad
cells
17 cells of which 9 are air flow
cells, 3 are static head cells
Cell height 13 cm
Cover
Nylon/PU, breathable, washable
cells Nylon/PU;
Max. load 36-150 kg
ENVIRONMENTAL PARAMETERS
Temperatures
Operation
10 - 35°C
Storage
-15 – 50°C
Shipment
-15 – 70°C
Relative humidity
Operation
20 – 80% => non-condensing
Storage
10 – 90% => non-condensing
Instructions for use DUO WAVE
EN
17