Humidifier performance
The following settings have been tested at 22°C ambient temperature:
Mask Pressure
RH output %
cm H
O (hPa)
2
Setting 4
4
85
10
85
20
85
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
Heated air tubing temperature cut-out: 41°C
Notes:
The manufacturer reserves the right to change these specifications without notice.
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
Do not use electrically conductive or antistatic air tubing.
The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Pressure sensor at air outlet:
Mask pressure
Flow derived values:
Leak
Value
1
Pressure measurement
:
2
Mask pressure
1
Flow and flow derived values
:
Flow
2
Leak
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
Pressure accuracy
Maximum static pressure variation at 10 cm H
Without humidification
With humidification
18
All manuals and user guides at all-guides.com
Setting 8
100
100
90
Range
4–20 cm H
0 (4–20 hPa)
2
0–120 L/min
Accuracy
±[0.5 cm H
0 (0.5 hPa) + 4% of measured value]
2
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
O (10 hPa) according to EN ISO 17510-1:2009
2
Standard air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
1
Nominal system output AH
Setting 4
Setting 8
6
>10
6
>10
6
>10
Length
Inner diameter
2 m
15 mm
1.8 m
15 mm
2 m
19 mm
Display resolution
0.1 cm H
O (0.1 hPa)
2
1 L/min
SlimLine air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
2
, BTPS