Procedure:
Switch on balance
Operate adjustment switch; for position see chap.14
The balance is now in service menu.
Press
Confirm with
displayed, e.g. 5000 kg
(example)
Select desired setting with
(example)
Put Ld is shortly displayed, the first adjustment point 0.000 kg is
displayed.
Unload the balance and confirm with
WAit, followed by Put ld is displayed, followed by the set
adjustment weight.
Place the adjustment weight, confirm with
The adjustment is carried out, after that the balance switches off
automatically.
(example)
An adjusting error or incorrect adjustment weight will generate an error message
("Err 4"), repeat the adjustment process.
TMBA_B-BAPS-e-2111
or
repeatedly until cAL E appears
, the adjustment weight selected as last is
.
or
and acknowledge by
.
33