Recommended Post-Processing Equipment:
Formlabs validated accessories
Formlabs validated washing units with approved solvent
Formlabs validated curing units
REQUIREMENTS
For biocompatibility compliance, IBT Resin was validated with a dedicated resin tank, build platform,
wash unit, and post-processing equipment that were not mixed with any other resins.
A. PRINTING
Printing guidelines and validated equipment can be found in the Manufacturing Guide at
support.formlabs.com.
B. WASHING
Washing guidelines and validated equipment can be found in the Manufacturing Guide.
C. POST-CURING
Post-Curing guidelines and validated equipment can be found in the Manufacturing Guide.
D. SUPPORT REMOVAL
Support marks can cause abrasion if not removed. See the Manufacturing Guide for recommendations.
E. CLEANING & DISINFECTION
1. Fully post-processed parts can be cleaned using a dedicated soft toothbrush with neutral soap
and room-temperature water. Do not use toothpaste or any other cleaning products on parts
printed with IBT Resin. After cleaning, always inspect parts for any cracks. Discard if any damage
or cracks are detected.
2. The appliances may be cleaned and disinfected according to facility protocols. Tested disinfection
method: soaking the finished appliance in fresh 70% IPA for 5 minutes. Do not leave the part in the
alcohol solution for longer than 5 minutes. After disinfection, inspect the part for damage or cracks
to ensure that the integrity of the designed part meets performance requirements.
F. STORAGE
1. Printed appliances should be stored in closed, opaque, or amber containers. Excess light exposure
over time may affect the color and performance of printed appliances.
2. Store cartridges at 10°C - 25°C (50°F - 77°F). Do not exceed 25°C (77°F) when in storage.
3. Keep cartridges closed & away from ignition sources.
G. DISPOSAL
1. Cured resin is non-hazardous and may be disposed of as regular waste.
2. Follow facility protocols for waste that may be considered biohazardous.
3. Unused liquid resin & solvents should be disposed of in accordance with local regulations.
4. Cartridge & contaminated packaging should be disposed of in accordance with local regulations.
H. ADVERSE EVENTS
To report any serious injury (temporary or serious permanent deterioration of health, a serious public
health threat, or death), contact Formlabs Inc. (+1 617 855 0762) or your local Competent Authority.