Specific warnings concerning use of the thermometer:
- Do not touch the protective glass around the thermometer's infrared lens.
- As the device has been pre-configured at the end of the manufacturing process, there is no need to calibrate it before first use.
Specific warnings concerning use of the tensiometer:
- Use an arm cuff size appropriate to the patient's arm to avoid inaccurate results.
- Do not remove the air tube connector.
- Avoid allowing any liquid to enter the air tube.
- It is recommended that you replace the arm cuff every year.
1.2. GENERAL PRECAUTIONS
- Only use this device in the situations specified and described by this user manual.
- This device is designed for external use.
- Use and storage: This device should always be stored in a clean, dry place. Refer to the technical characteristics (air temperature range, relative humidity,
atmospheric pressure).
- Do not expose this device to extremes of temperature or to the sun.
- To obtain reliable and stable results, it is recommended that whenever changing to an environment with a significantly different air temperature you leave
the device at this temperature for 15-20 minutes before use.
- Do not submerge the device in any form of liquid. Do not use this device if it has been submerged in a liquid.
- Do not expose this device to electrical shocks.
- Avoid electrostatic charge build-up.
- Never drop the device.
- Follow the maintenance instructions indicated in this user manual.
- This product has no components that can be repaired by the user. Any unauthorised modification made to the product voids the guarantee, as well as any
applicable regulatory approval or certification.
- Do not insert any object into the openings of the device. Do not use any sharp object when using the device.
- Only use accessories and spare parts specified by the manufacturer or sold with the device.
- Do not use if the device or its accessories are damaged, or in the event of an anomaly or malfunction.
- Keep this device and its accessories out of the reach of children. Keep small components that could be swallowed away from children.
- This device is not designed to be used by people (including children) with reduced physical, sensory or mental capabilities, or by inexperienced people or
those lacking the necessary knowledge, except if they use the device in the presence of someone responsible for their safety; under supervision or following
the operating instructions of the device.
- This device is designed to be used in an electromagnetic environment where radiated RF disturbances are controlled. The user of the device can help to avoid
electromagnetic interference by ensuring a minimum distance between mobile and portable RF communication equipment (transmitters) and the device; as
recommended at the end of this guide, which depends on the maximum output power of the communication equipment.
- Readings can be distorted if the device is used near a television, microwave oven, mobile telephone or any other device with an electric field.
1.3. CYBERSECURITY INFORMATION
When used locally, personal data is stored and encrypted with 256-bit AES on the device.
When used in connected mode, personal data is transferred and stored by an approved health data hosting service, in compliance with a secure protocol.
To ensure the security of your data, do not share your password with another person. Do not leave your device unlocked and/or unattended.
Checks for updates are automatically made when the device is started up. It is recommended that you always update your Visiocheck® software.
2. INTENDED USE
Visiocheck® BW-X07HD is intended for use in health care establishments, medical practices, at home, and in other health care sectors. This device is designed
for monitoring vital signs (temperature, heart rhythm, blood pressure, oxygen saturation and pulse rate); it can be used by healthcare professionals or by
people who have been trained to use this device.
Indication:
- Oximeter: All categories of patients (adults, children, infants)
- Thermometer: All categories of patients (adults, children, infants)
- Tensiometer: Adults, Children
26
BewellConnect - VisioCheck - BW-X07HD - User Manual - 2018 - 04 - V1
All manuals and user guides at all-guides.com
- ECG: Adults
This device enables communication via a video conferencing system. It cannot be used as a telephone.
3. WORKING PRINCIPLE
Visiocheck® BW-X07HD is fitted with in-built TEMP, ECG, SpO
, BPM modules, which are intended to take physiological measurements, and perform additional
2
functions.
The ECG module enables an electrocardiogram to be produced (7 leads: AVR, AVL, AVF, I, II, III, V). When the electrodes are placed on the skin, the device
records the potential difference resulting from electrical waves spreading through the body. This enables the heart rate to be recorded, as well as the nature
of conduction and heart rhythm.
The BPM module (tensiometer) enables arterial blood pressure and heart rate to be recorded using the oscillometric method, which is automated and non-
invasive. The air pump allows the upper arm cuff to be automatically inflated. By measuring the pulse when the arm cuff is inflated, an algorithm is able to
calculate, based on the sensor reading, systolic arterial pressure, diastolic pressure and heart rate.
The TEMP module (temperature) uses an infrared sensor to measure heat radiated by the body from the forehead or temple. By making a comparison with the
surrounding temperature, an algorithm enables core body temperature to be calculated.
The SpO
module (oxygen saturation measured with pulse oximetry) is a non-invasive test that enables levels of oxygen carried by haemoglobin in the
2
capillaries to be measured. It enables oxygen saturation, pulse rate and perfusion index to be calculated.
Using an Android system, Visiocheck® includes functions such as a user interface, data storage and transmission (Bluetooth, 3G, 4G, WiFi), video conferencing,
etc.
4. GENERAL INFORMATION
4.1. CHARGING THE BATTERY
Before using for the first time, fully charge the device for around 4 hours.
When the symbol "battery" appears, recharge the device.
1. Connect the cable (micro USB part) to the designated port on the device.
2. Connect the USB cable to the charger, connect the Zone adaptor to the charger, then to the mains supply.
3. When charging, the battery indicator is blue, then turns green when the battery is fully charged.
4. When charging is complete, disconnect the device.
Notes:
- Before using the device, check the battery level. A low battery (less than 15%) can jeopardise the reading being taken.
- In the Settings menu, the battery level is displayed (Status bar).
- Only use the charger and accessories provided.
4.2. TURNING ON AND OFF
Turning on: Press and hold the ON/OFF button until the screen lights up.
Turning off: Press and hold the ON/OFF button until the screen turns off.
Putting the device into standby: Briefly press the ON/OFF button.
4.3. MODULE INSPECTION
When starting up the device, the modules are automatically tested.
Success: The module is working correctly.
Fail: The module has a fault. Restart the device. If the problem persists, contact customer service.
BewellConnect - VisioCheck - BW-X07HD - User Manual - 2018 - 04 - V1
27