Chapter 1: Ultravision™ – overview and general features
This section describes the indications for use for the Ultravision surgical smoke visual field clearing system and its features.
Caution
Read all warnings, cautions, and instructions provided with the Ultravision system before use. Failure to do so could result in
death or serious injury to the patient or operator.
Read the following associated individual manuals as these contain safety relevant information.
Part No.
DLU-001-025
DLU-001-026
DLU-001-004
DLU-001-005
DLU-001-009
DLU-003-004
DLU-004-001

Indications for use

The Ultravision system is indicated for the clearance of smoke and other particulate matter that is created during laparoscop-
ic surgery.
Warning
Ultravision is not intended to be used in non-laparoscopic surgery procedures.
Do not allow the Ionwand to come in to contact with the uninsulated portion of an electrosurgical instrument during opera-
tion. In such a situation the Ultravision generator issues a continual audio-visual proximity indication. Continued contact for
in excess of 10 minutes may result in an electrical discharge from the electrosurgical instrument that may be sufficient to
cause atrial fibrillation depending on proximity to the thoracic cavity.
Use Ultravision with caution in the presence of internal or external pacemakers. Similar to electrosurgical devices,
Ultravision can affect the performance of a pacemaker. Consult the pacemaker manufacturer or hospital Cardiology
Department for further information when use of Ultravision is planned in patients with cardiac pacemakers.
Similar to electrosurgical devices, Ultravision may also interfere with the performance of an internal cardiac defibrillator
(ICD). If the patient has an ICD contact the ICD manufacturer for instructions before performing an electrosurgical
procedure with Ultravision.

Contraindication

None.
Parts shipped with the Ultravision™ system
The Ultravision system comprises both reusable and single-use, sterile disposable items. The single-use, sterile dispos-
able items are supplied separately. Furthermore, the Mains Converter is an optional form of powering the generator that is
supplied separately or within an alternate Ultravision system starter pack variant. Contact your local sales representative in
order to establish a supply of these items. All other items required for Ultravision to function are shipped with the Ultravision
system.
Before use, confirm that the shipping carton contains the following items:
1. Generator
2. Battery x 2 (battery powered system only)
3. Recharging station (battery powered system only)
4. Mains Power supply unit for recharging station (battery powered system only)
5. Mains converter including power supply (mains powered system only)
6. Solid-plate patient return adaptor
7. Split-plate patient return adaptor
8. User manual
6
Manual Description
Ionwand Pack IFU
Patient Return Adaptor IFU
Battery Recharging Station IFU
Battery IFU
Ultravision User Guide
Ultravision 5mm Trocar IFU
Mains Converter IFU
x1
x2
x1
x1
x1
x1
x1
x1