Electromagnetic emissions – Ultravision Battery Recharging Station
The Ultravision battery recharging station is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Ultravision battery recharging station should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environ-
ment – guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Electromagnetic immunity – Ultravision Battery Recharging Station
The Ultravision battery recharging station is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Ultravision battery recharging station should assure that it is used in such an environment.
I
test
MMUNITY
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
NOTE
is the a.c. mains voltage prior to application of the test level.
U
T
32
Compli-
Electromagnetic environment - guidance
ance
Group 1
The Ultravision battery recharging station uses RF energy only for its inter-
nal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment. RF emissions
The Ultravision battery recharging station is suitable for use in all estab-
Class A
lishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
Class A
purposes.
Complies
IEC 60601test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
100kHz repetition frequency
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0 % U
; 250/300 cycle
T
30 A/m
Compliance level
Electromagnetic environment
± 8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
± 15 kV air
synthetic material, the relative humidity
should be at least 30 %.
± 2 kV for power
Mains power quality should be that of a
supply lines
typical commercial or hospital environ-
ment.
± 1 kV for input/
output lines
± 1 kV line(s) to line(s)
Mains power quality should be that of a
typical commercial or hospital environ-
± 2 kV line(s) to earth
ment.
0 % U
; 250/300 cycle
Mains power quality should be that of a
T
typical commercial or hospital environ-
ment. If the user of the Ultravision battery
recharging station requires continued oper-
ation during power mains interruptions, it is
recommended that the Ultravision battery
recharging station be powered from an
uninterruptible power supply or a battery.
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
guidance