Télécharger Imprimer la page

Sanitas SEM 40 Mode D'emploi page 18

Masquer les pouces Voir aussi pour SEM 40:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
• In case of known or acute cardiac arrhythmias and other disorders in
stimulus formation and conduction of the heart
• If suffering from attacks (e.g. epilepsy)
• If pregnant
• In case of cancer
• After operations where increased muscle contractions might impair the
healing process
• Never use near the heart. Stimulation electrodes should
never be placed anywhere on the front of the thorax
(marked by ribs and breastbone), but above all not on the
two large pectoral muscles. Here it can increase the risk of
ventricular fibrillation and lead to cardiac arrest
• On the bony part of the cranium, near the mouth, throat or
larynx
• Near the throat / carotid artery
• Near the genitals
• On acutely or chronically diseased (injured or inflamed) skin (e.g. in the
event of painful and painless inflammation, reddening, skin rashes (e.g.
allergies), burns, bruises, swellings, open wounds and wounds in the
process of healing, on operation scars in the process of healing)
• In environments with high humidity such as in the bathroom or when
having a bath or shower
• Do not use after consuming alcohol
• When a high frequency surgical device is connected at the same time
Before using the machine, you should consult with the doctor in charge
of your treatment in the event of the following:
• Acute diseases, especially if hypertension is suspected or actually exists,
or blood-clotting disorders, tendency to thromboembolic diseases or in
case of malignant new masses
• All skin diseases
• Chronic pain disorders which have not been clarified, regardless of the
region of the body
• Diabetes
• All sensitivity disorders with reduced pain awareness (such as metabolic
disturbances)
• Medical treatments carried out simultaneously
All manuals and user guides at all-guides.com
• Complaints arising as a result of the stimulation treatment
• Constant skin irritation as a result of prolonged stimulation at the same
IMPORTANT!
Use the Digital EMS/TENS solely:
• On humans.
• For the purpose for which it was developed and in the manner specified in
• For external application.
• With the original accessory parts which are supplied and can be re-
PRECAUTIONARY MEASURES:
• Always remove the electrodes from the skin with a moderate pull in order
• Keep the machine away from heat sources and never use it near (~1 m)
• Do not expose the machine to direct sunlight or high temperatures.
• Protect the machine from dust, dirt and humidity. Never submerge the
• The machine is suitable for private use.
• For reasons of hygiene, the electrodes should only be used on one person.
• If the machine is not functioning properly, or if you start feeling unwell or
• To remove or move the electrodes, first switch off the machine or the ap-
• Never modify any electrodes (e.g. by cutting). This leads to a higher
• Do not use when asleep, when driving a vehicle or at the same time as
• Never use with any activities where an unforeseen reaction (e.g. increased
18
electrode point
these instructions for use. Any improper use can be hazardous.
ordered, otherwise the warranty becomes null and void.
to avoid injury in the event of highly sensitive skin.
short-wave or microwave equipment (e.g. mobile phones), as this may
lead to unpleasant current spikes.
device in water or other liquids.
there is any pain, stop the application immediately.
propriate channel in order to avoid unwanted irritation.
current density and can be dangerous (max. recommended output value
for electrodes: 9 mA/cm
, an effective current density over 2 mA/cm²
2
requires greater attention).
operating machinery.
muscle contraction despite low intensity) may be dangerous.

Publicité

loading